COMMERCE BUSINESS DAILY ISSUE OF JULY 6,1999 PSA#2381USDA, ARS, NPA, 1201 Oakridge Drive, Suite 150, Fort Collins, CO
80525-5562 66 -- WALK-IN GROWTH CHAMBER SOL RFQ-021-82HW-99 DUE 080699 POC
Contact Point, Fax to Lee Shea, Contract Specialist A Walk-In Growth
Chamber is required by the National Seed Storage Laboratory, Fort
Collins, CO. The space available measures 18 Feet 8 Inches X 8 Feet X
10 Feet High with minamal air conditioning in the area. Room shall be
self contained with insulated walls and floor. Flooring shall be water
tight. Interior dimensions shall be a minimum of 6 Feet (L)X 6 Feet
(W) X 7 Feet (H) with growth area a minimum of 36 feet(squared).
Temperature range of 5 degrees C to 40 degrees C, Temperature control
of +/-1.0 degrees C and Temperature uniformity of +/-2.0 degrees C.
Light intensity shall be a minimum of 5,000 foot candles or 54,000 LUX
(at least half of full sun). Installation is included according to
installation plan provided by ARS. Posted 07/01/99 (W-SN349401).
(0182) Loren Data Corp. http://www.ld.com (SYN# 0392 19990706\66-0021.SOL)