COMMERCE BUSINESS DAILY ISSUE OF JULY 6,1999 PSA#2381USDA, ARS, NPA, 1201 Oakridge Drive, Suite 150, Fort Collins, CO
80525-5562 66 -- WALK-IN GROWTH CHAMBER SOL RFQ-021-82HW-99 DUE 080699 POC
Contact Point, Fax to Lee Shea, Contract Specialist, (970)229-5531 A
Walk-In Growth Chamber is required by the National Seed Storage
Laboratory, Fort Collins, CO. The space available measures 18 Feet 8
Inches X 8 Feet X 10 Feet High with minamal air conditioning in the
area. Room shall be self contained with insulated walls and floor.
Flooring shall be water tight. Interior dimensions shall be a minimum
of 6 feet (L) X 6 feet (W) X 7 feet (H) with growth area a minimum of
36 feet (squared). Temperature range of 5 degrees C to 40 degrees C,
Temperature control of +/-1.0 degrees C and Temperature uniformity of
+/-2.0 degrees C. Light intensity shall be a minimum of 5,000 foot
candles or 54,000 LUX (at least half of full sun). Installation is
included according to installation plan provided by ARS. Posted
07/01/99 (W-SN349744). (0182) Loren Data Corp. http://www.ld.com (SYN# 0386 19990706\66-0015.SOL)