|
COMMERCE BUSINESS DAILY ISSUE OF FEBRUARY 25,1999 PSA#2290AFMETCAL Det 1/MLK, 813 Irving-Wick Dr W, Suite 4M, Heath, OH
43056-6116 66 -- PHOTOMETRIC CALIBRATION SYSTEM SOL F33660-99-R7007 DUE 041499
POC Buyer: Janet Casler, 740-788-5041, Contracting Officer: Glenna
Pound, 740-788-5043 The Air Force Metrology and Calibration (AFMETCAL)
program requires 5 each Photometric Calibration Systems in accordance
with Directorate of Metrology Purchase Description PD99MLEEJG1-1,
dated 11 Dec 98. Requirement includes a 5 year worldwide warranty and
data with each unit and an Option for 0-5 each in FY00. General
Description: The iphotometric calibration system is a system of light
sources and detectors for calibrating a wide variety of luminance and
illuminance meters. The system may be comprised of a source for
luminance and a separate source for illuminance, or they may be
combined into one source with accessories to produce fields of light
for calibrating luminance and illuminance meters. The power supplies
and controllers may also be combined into one unit or may be separate
items for the luminance and illuminance sources. The photometric
calibration system shall include standard detectors to be used to
transfer calibration from the Air Force Primary Standards Laboratory to
the precision measurement equipment laboratories. The luminance source
shall be a sphere type source to provide a uniform field of light at
the output port for calibration of luminance meters. The diameter of
the output port shall be from three to four inches. The color
temperature of the luminance source shall be Illuminant A (2856K
 50K) and shall be adjustable to 3100K. The spectral output of
the source must be planckian in nature over the range of 350 to 800 nm
with no dips or spikes in the wavelength spectrum. The luminance
produced shall be from 0 to 20,000 fL (footlamberts) (68,000
candela/square meter) when the color temperature is set to 2856K. The
luminance levels shall be adjustable over the full luminance range to
be able to set at least four luminance levels in each decade over the
range of 10 to 20,000 fL. The color temperature shall remain constant
when setting luminance levels. The luminance levels shall not be set by
changing the lamp current or temperature settings. However, the
controller may adjust the lamp current to maintain a constant luminance
value while maintaining the color temperature within + or -- 50K of
2856K. The luminance shall be uniform to better than + or -- 2% across
the diameter of the output port. The output port shall have a frame
capable of mounting various attachments such as opal glass diffusers,
color filters, and detector holders. The luminance output level shall
be controlled by utilizing a photopically corrected detector mounted on
the sphere wall to stabilize the luminance levels to + or -- 0.5% or
better. The responsivity of the detector shall not change as the sphere
warms up when heated by the lamps. The illuminance source shall provide
a collimated field of uniform light for calibration of illuminance
meters. The collimated output of the illuminance source shall give a
spot of at least two inches diameter where the illuminance is uniform
to better than + or -- 2%. The color temperature of the illuminance
source shall be Illuminant A (2856K + or -- 50K)and shall be adjustable
to 3100K. The spectral output of the source shall be planckian in
nature over the range of 350 to 800 nm with no dips or spikes in the
wavelength spectrum. The illuminance levels shall be from 0 to 2000
footcandles (fC) when the source color temperature is set to 2856K. The
illuminance levels shall be adjustable over the full illuminance range
to be able to set at least four illuminance levels in each decade over
the range of 1 to 2,000 fC. The color temperature shall remain constant
at all illuminance levels. The illuminance levels shall not be adjusted
by changing the lamp current or temperature settings. However, the
controller may adjust the lamp current to maintain a constant luminance
value while maintaining the color temperature within + or -- 50K of
2856K. The output optics system shall be capable of inserting
attenuation and color filters. The output beam shall be collimated such
that the effective photometric distance from the lens shall appear to
be offset by two meters or more when the illuminance meters are set in
the position of the highest illuminance levels. The calibration system
shall include a standard detector for luminance and illuminance. The
luminance detector shall be in addition to the luminance monitor
detector mounted in the luminance sphere source. These standard
detectors shall be used to transfer calibration from the Air Force
Primary Standards Laboratory to the precision measurement equipment
laboratories worldwide. The spectral response of the luminance detector
shall match the V( ) photopic spectral response curve with a value of
F1' of 2% or less, as calculated according to CIE publication No. 69
(1987). The luminance detector shall have a field of view of 3 degrees
or less. The optics of the luminance detector shall focus on the exit
port of the luminance source within the length of the included rail
system. The entrance aperture diameter of the luminance meter shall not
be larger than 1.5 inches. The detector shall be capable of being
mounted along the optical axis of the luminance source. The illuminance
detector shall have a response at 30 degrees incidence angle of within
+ or -- 2% of the ideal cosine response. The spectral response of the
illuminance detector shall match the V( ) photopic spectral response
curve with a value of F1' of 2% or less, as calculated according to CIE
publication No. 69 (1987). The diameter of the sensitive area of the
illuminance detector shall not be greater than one inch. The detector
shall be capable of being mounted along the optical axis of the
illuminance source. The redout unit for both the luminance and
illuminance detectors shall be able to display the readings in either
footcandles or lux (lumens per square meter) for illuminance and in
either footlamberts or candela per square meter for luminance. The
calibration system shall include all necessary power supplies,
controllers, and mounting system to operate the light sources and
perform calibrations with the luminance and illuminance detectors. The
sphere source controller shall display a reading of the luminance
value from the controller detector in units of footlamberts or candela
per square meter. The units of the display shall be selectable by the
user. The calibration of the luminance readout shall be adjustable in
a manner that will permit recalibration of the luminance levels by
comparison to a standard luminance meter. The controllers for all the
sources shall turn on the lamps by slowly ramping up the current to
prevent damage to the lamps by thermal shock. Both the luminance and
illuminance sources shall include a rail and holder system that can be
used to read the distance from the source to the standard and test
instrument detectors and keep the center of the light field centered on
the standard and test meters. The rail shall not be longer than 1.5
meters. The input power requirements are: 120 Vrms +/- 10%, 50 to 60
Hz. Instrument shall include a power cord with a minimum length of 6
feet. The mechanical and physical requirements are as follows:
Commercial parts and materials that are used shall have properties
consistent with a precision laboratory instrument. The equipment shall
be designed and constructed of materials that meet the requirements of
this specification and in accordance with the best commercial practices
for industry approved equipment. It shall be designed and manufactured
to maximize reliability, maintainability, accuracy of operation, and
ease of operation. All parts such as shafts, bearings, regulators,
switches, controls, etc., shall have proper clearance and adjustments.
They shall work together so that the equipment will supply the rated
requirements without unnecessary strains, vibrations, or overheating.
They shall be able to withstand the conditions met in shipping,
storage, installation, and service. All materials and parts of this
equipment shall be capable of withstanding the environmental
requirements of this specification. Electrical, electronic, and
electromechanical items shall be capable of operating in an ambient
temperature of 0 to 50 deg C without loss of required performance. All
accessories necessary for normal operation of the equipment shall be
supplied. The unit shall be assigned a model number designation by the
contractor. Each individual unit shall be assigned a unique serial
number designation by the contractor. A complete list of the
performance/functional requirements is contained in the purchase
description attached to the solicitation package. Solicitation Number:
F33660-99-R7007. FOB Destination. Delivery to be a minimum of 3 each
within 180 calendar days after contract award and continue at a minimum
rate of 3 each every 30 days thereafter until complete. All
solicitation requests must be in writing. Written requests can be faxed
to 740-788-5157. Oral requests will not be accepted. Request for
proposal will be issued approx. 15 Mar 99. Closing date will be approx.
14 Apr 99. All responsible sources may submit a proposal, which shall
be considered. Special note to all prospective offerors: DOD will not
awardcontracts to contractors that have not registered in the Central
Contractor Registration (CCR) data base. Registration may be done by
assessing the CR via the Internet at http://ccr.edi.disa.mil or call
the DOD electronic Commerce Information Center at 1-800-334-3414.
Posted 02/23/99 (W-SN301237). (0054) Loren Data Corp. http://www.ld.com (SYN# 0380 19990225\66-0012.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|