COMMERCE BUSINESS DAILY ISSUE OF FEBRUARY 23,1999 PSA#2288National Institutes of Health, National Heart, Lung, and Blood
Institute, Contracts Operations Branch, 6701 Rockledge Drive, Room
6100, MSC 7902, Bethesda, MD 20892 66 -- MODEL 1.0 REAL-TIME VOLUMETRIC SCANNER SOL NHLBI-HL-P-99-383 POC
Gina M. Cianflone, Contract Specialist, (301) 435-0363; Debra C.
Hawkins, Contracting Officer, (301) 435-0366 The National Heart, Lung,
and Blood Institute intends to purchase the brand name or equivalent
Model 1.0 Real-time Volumetric Scanner for real-time three dimensional
imaging of the heart manufactured by 3-D Ultrasound Inc., 700 West
Main St., Durham, NC 27701. Specifically, the contractor shall deliver
and install the Model 1.0 Volumetric Scanner for real-time
three-dimensional echocardiography or equal to the National Heart,
Lung, and Blood Institute. The system shall have all of the following
mandatory minimum specifications: (1) It shall be an independent system
that is portable. (2) The Model 1.0 shall include the following:
TRANSDUCERS: (a) 2000 element 2.5 MHZ active matrix array, (b) 2000
element 3.5 MHZ active matrix array, (c) 2000 element 3.5 MHZ active
matrix array. BEAMFORMER: (a) 256 channels 16:1 parallel processor
(4096 ultrasound signal processing channels). DISPLAY: (a) Conventional
2D real-time display, (b) Multiple real-time simultaneous planes with
two possible formats, (c) Two Orthogonal B-Scan planes with 2 C-Scan
planes in Orthographic Projection. The B-Scan and the C-Scan planes are
adjustable, (d) Two orthogonal B-Scan planes with 2 C-Scan planes in
Sheared Projection. The B-Scan planes and the C-Scan planes shall be
adjustable. VOLUME STORAGE: (a) 256 Mbytes on line Digital Storage (six
seconds of Cardiac Date). It shall have the ability to be separated
into four 1.5 second segments, (b) Magneto Optical Disk Drive for
saving studies to disk. Sony Model; SMO-F541/SMO-F544, (c) Data
management and review software, (d) Measurement Software, (e)
ECG-3-lead with separate ECG output. MONITOR: (a) Sony Trinitron Color
Graphic Monitor, model PGM-100PIMD with tilt and swivel. COLOR FLOW
AND DOPPLER: (a) Doppler modes on 2.5MHz matrix array, (b) Choice of
velocity or frequency display with angle correction, (c) PRF selection,
(d) ECG display, (e) Readout of frame rate, (f) Adjustable gain, (g)
Wall filter selection, (h) Baseline shift, (I) Color volume sector
volume and position adjustments, (j) Selectable threshold for color
flow, (k) Color map selection, (l) Color reject. (3) The offeror shall
deliver and install the system to the site furnished by NIH, and shall
provide immediate on-site training. (4) The offeror shall provide
information on warranty and repair. A firm-fixed price type contract
will be negotiated. Delivery and installation shall occur within thirty
(30) calendar days after contract award. Tests will be performed to
ascertain that the equipment is functional before final acceptance.
This is a 100% Small Business Set-Aside. The applicable Standard
Industrial Classification Code (SIC) is 3826, size standard 500
employees. The proposals may be due ten (10) calendar days from the
publication date of this synopsis. Interested parties should submit a
written request (via mail or facsimile transmission to [301] 480-3345)
for a copy of the solicitation number NHLBI-HL-P-99-383 to the
National Heart, Lung, and Blood Institute, Contracts Operations Branch,
Procurement Section, BuildingRKL2, Room 6150A, 6701 Rockledge Drive,
Bethesda, Maryland 20892-7902. Attention: Gina M. Cianflone. All
responsible sources may submit a proposal, which if timely received
shall be considered by the agency. See numbered Note(s) 12. Posted
02/19/99 (W-SN300263). (0050) Loren Data Corp. http://www.ld.com (SYN# 0328 19990223\66-0009.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|