COMMERCE BUSINESS DAILY ISSUE OF OCTOBER 29,1998 PSA#2211Naval Research Laboratory, 4555 Overlook Ave. S.W., Washington, D.C.,
20375 66 -- OPTICAL SPECTRUM ANALYZER SOL N00173-99-Q-0006 POC Darlene Copp,
Purchasing Agent, (202) 767-2374 The Naval Research Laboratory has a
requirement for an Optical Spectrum Analyzer with 600-1750nm wavelength
range, +/- 0.05nm (1510-1580nm) wavelength accuracy, +/- 0.01nm
(1510-1580nm) wavelength linearity, 0.015nm resolution, 60dB dynamic
range (at peak +/- 0.2nm), 70 dB (+/- 0.4nm), 0.3dB level accuracy, +/-
0.05dB level linearity, +/- 0.1dB (1510-1580nm) level flatness, +/-
0.05dB polarity dependance (at 1300/1550nm), built-in wavelength
standard, < 60 lbs in weight, internal 3.5 floppy disk drive,
standard DOS format, built-in printer, GPIB interface, with color
display, part #AQ6317 or equal. Please visit our website
http://heron.nrl.navy.mil for copies of solicitation package. Paper
copies will not be issued. See Note(s): 1. Posted 10/27/98
(I-SN265880). (0300) Loren Data Corp. http://www.ld.com (SYN# 0252 19981029\66-0002.SOL)