COMMERCE BUSINESS DAILY ISSUE OF APRIL 28,1998 PSA#2083Arnold Engineering Development Center, Attn: AEDC PKP, 100 Kindel
Drive, Suite A332, Arnold AFB, TN 37389-1332 66 -- VACUUM CALIBRATION SYSTEM SOL F40650-98-R-0019 DUE 070998 POC
Alan Fudge, Contracting Officer, 931-454-4406 E-MAIL:
fudge@hap.arnold.af.mil, fudge@hap.arnold.af.mil. Provide a vacuum
calibration system designed primarily for the evaluation, acceptance
test, and routine calibration of various types of vacuum gauges over
the range of 1 x 10 to the -7th to 1000 Torr. The system shall be
computer controlled and automated to read the standard gauges and the
gauges being calibrated. The maximum total uncertainty in the system
shall not exceed plus or minus 0.05% of indicated pressure and
temperature coefficients. The standard gauges in the system shall be
calibrated by or traceable to the National Institute of Standards and
Technology (NIST). The contract performance period will be
approximately 270 days. Delivery and installation will be at Arnold
AFB, TN. Firms responding to this synopsis should do so in writing and
indicate whether or not they are a minority business concern. Requests
for the solicitation documents may be faxed to 931-454-7330. (0114) Loren Data Corp. http://www.ld.com (SYN# 0289 19980428\66-0004.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|