Lewis.Fran@epamail.epa.gov. DESC: Portable GC/MSD Standard System
w/dual MPV, fixed & removable traps/Membrane/Mass flow meter Cryo
cooling/Flat panel monitor/16MB Ram/Softpack/Hellium Bottle Field Kit
w/o libraries. 1/s Dual-stage Turbomolecular, w/Diaphragm Roughing
Vacumn Pump/Purge&Trap Single Sparger&Software/Purge&Trap Single
Sparger & Software/Spectratrak NIST Mass Spectra Library Bundle,
(000-411 and 415).16Mb DRAM installed & tested on SBC(Factory installed
only) Internal Modem&Remote Control Software (28.8KB,V.34)(Factory
installed only). PCMIA Type I-III 3.5" Card socket/Controller. PCMCIA
260MB Removable hard Drive. Field Operations Maintenance & Supply
Kit/Softpak Instrument Transport case/Softpack Roughing Pump/Accessory
transport case/InstrumentHard Sided Transport case/Roughing
Pump/Accessory Hard Sided Transport case/Custom Commercial Electrical
Generator, 2500 Watts, 110 Volts/Three Day On-Site Installation and
Training(US ONLY) 9 month Warranty Extension following 3 month warranty
with upgraded service providing on-site repair service within Zone
B/Custom CTC A200SE GC Autosampler with 100 position rack/Custom
Mounting Hardware, Cabling and Software for CTC A200SE Autosampler,
Custom External Device Interface Kit. Interface Kit for Tekmar 3000
includes injection port and software/Custom Tekmar interfacing.One Day
service engineer time for interfacing SpectraTrak to Tekmar system/HP
Wiley Library of 138,000 Mass Spectra(PBM Format,84Mb disc
space)/custom hardDrive upgrade,2.1GB or better/Portable
chemstation(computer)w/remote control software and cable.HP5000 laptop
computer-Pentium based 90 mhz with 16MB RAM,810 MB Hard Drive, TFT
Color Display and 28.8 Kbaud PCMCIA Modem, or equivalent,Syquest
Transportable Drive w/parallel/VSpectraScan Interface. HP Portable
Deskjet Laser 3 40C. Design Qualifications:The following analysis will
be required: Ambient air toxics-detection level of 1 ppb/v for EPA air
toxics with at least 1L of air sample collected on absorbent.Ambient
air toxics-Direct injections of 1mL for emergencies and soil gas
analyses.VOA-Analyze 5mL of water by purge&trap.VOA-Sparging
capability.BNA-Spitless injections 1-3 uL of methylene chloride&hexane
extracts.BNA-autosmapler available.Operating Environment:Must
becapable of operating from field generator for power.Temperature
range-20to95F.Humidity up to 100%.Must be rugged to transported in the
back of a truck over a dirt road and not lose vacuum,tuning,and
calibration. Performance (capability):Ambient air toxics-Interfacing
with a PE or Entch preconcentrator for canister
analysies.VOA-Interfacing with Tekmar ALS. BNA-Large volume injection
capability-50uL(thermal concentration).GC:Heated:ambient to 300Cwith
multiramp programing.CO2 capability for sub ambient cooling.Large
enough for a 60m capillary column.Handle 0.2 to 0.5mm capillary columns
with flows 1-5 ml/min.Splitless injector EPC autosampler
capabilities.MS:Range 35-600amu.Sensitivity 5NG of DFTPP&5NG of BFB.
Must have SIM mode.Scan speed 0.5 to 1 secfor 35-450amu. Pump down in
less than 1 hour to operating condition.Vacumn in 10-5 torr range.Must
have an isolation valve. Software:Windows 95OS. Mouse
controlled(compatability with HP Enviroquant is a plus)Automatic Tune
check routines for BFB and DFTPP.Automatic Initial&Continuous
Calibration check routines for %RSD&%D. Automatic TIC library search
routine.Air Toxics:The absorbent tubes must be transportable to collect
the samples apart from the instrument and then installed for
analyses.Desorption temp at least 200 C. IN EMERGENCIES,MUST BEABLE TO
HAVE FIELD ENGINEER ON SITE WITHIN 24 HR OF INITIAL CALL. THREE HR
TELEPHONE RESPONSE CALL TO TECHNICAL QUESTIONS 8-5 EST. MUST HAVE TEST
RESULTS(QC)FROM THE MANUFACTURER AS TO THE OPERATIONS BEFORE
SHIPPING.All necessary manuals supplied. Initial start-up electronic
self tests and software validation:Back up copies of software provided.
Required Acceptance tests to confirm the design qualifications:In House
Initial Installation using portable generator:Tuning(DFTPP&BFG)at 5ng
injection-Must pass EPA's CLP criteria.Initial Calibration-Must have
less than 20%RDS,CLP criteria for VOA&BNA.Continous
Calibration&Tuning-Must have less than 25%Dover a 10 hr period of
operation.QC samples-Must-be within 25% of the mean accepted value for
the VOA and NIST air toxic sample at 1 ppb.Area counts for IS for
VOA,air toxics,&BNAs-50 to 150 of mean initial calibration.Field
Demonstration for VOA or air toxics. Instrument must MAINTAIN TUNE OVER
A 10 HR PERIOD(DFTPP&BFB)MAINTAIN INITIAL CALIBRATION OVER A 10HR
PERIOD FOR AIR,VOA&BNA ANALYSES.***** (0210) Loren Data Corp. http://www.ld.com (SYN# 0364 19970731\66-0014.SOL)