|
COMMERCE BUSINESS DAILY ISSUE OF MAY 28,1997 PSA#1854National Institute of Standards & Technology, Acquisition & Assistance
Div., Bldg. 301, Rm B117, Gaithersburg, MD 20899 66 -- TUNABLE DIODE LASER SYSTEM SOL 53SBNB7C1119 DUE 053097 POC Julie
Liu (301)975-6727, Fax (301)963-7732 WEB: NIST Contracts Homepage,
http://www.nist.gov/admin/od/contract/contract.htm. E-MAIL: NIST
Contracts Office, Contract@nist.gov. This announcement modifies the
original announcement published April 14, 1997 and the modification
published May 5, 1997. The date and time for receipt of quotations is
extended to 3 PM EST on May 30, 1997. The specification stating, "at
least one scheme shall be capable of measuring a minimum CO absorbance
of 10E-5 with a signal-to-noise ratio of unity, with an averaging time
of no more than 1 second." is replaced by, "at least one scheme shall
be capable of measuring a minimum CO absorbance of 0.00001 with a
signal-to-noise ratio of unity, with an averaging time of no more than
1 second." The following questions have been received from potential
offerors: 1) Q. A GRIN lens is specified for the fiber collimator
output of the single mode fiber. If an anti-reflection coated aspheric
lens provides equivalent operation, can this be substituted? A. An
anti-reflection-coated aspheric lens may be substituted in place of the
GRIN lens at the output of the single mode fiber if it provides
equivalent operation.2) Q. The output laser power from the
optically-isolated fiber is specified as at least 5mW. Is it possible
to substitute an otherwise suitable laser source providing at least 2
mW of power from the fiber output? A. The output laser power from the
optically-isolated fiber is specified as at least 5 mW for the purpose
of providing adequate power to survive attenuation by soot. For the
laser wavelength of 1.55 microns along a 2 cm path, typical soot volume
fractions in the vicinity of a fire could attenuate as much as 90% of
the incident beam intensity. In light of this, substitution of a laser
with 2 mW usable power for the specified 5 mW laser is acceptable only
if the system is capable of measuring the required CO absorbance of
0.00001 with low signal levels resulting from soot attenuation. 3) Q.
Detection scheme [1] is described as an electronic balancing method and
must be capable of absorbance sensitivity of 10 -5. The reference type
for this circuit described in reference 1 is not electronically
balanced and has only been shown to allow absorbance sensitivity of 10
-3. Is an electronically balanced direct absorption detector such as
described by P.C.D. Hobbs, "Ultrasensitive laser measurements without
tears, " Applied Optics, Vol. 36, No. 4, pp. 903-920, 1997 acceptable?
A. Yes, this is acceptable for detection scheme [1]. 4) Q. Will the
Government accept alternative approaches to the specified system? A.
Offerors are encouraged to submit alternate approaches. First, submit
a quotation on the system as specified, then submit a separate quote
for the alternate. (0143) Loren Data Corp. http://www.ld.com (SYN# 0270 19970528\66-0001.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|