COMMERCE BUSINESS DAILY ISSUE OF APRIL 8,1997 PSA#1819National Cancer Institute, Research Contracts Branch, PSCS, 6120
Executive Blvd, Eps/room 638, Bethesda, MD 20892-7227 66 -- WALLAC 1420 VICTOR MULTILABLE COUNTER SOL RFQ70068-NS DUE 042897
POC Patricia Haun, Contracting Officer, 301-496-8613 The National
Cancer Institute (NCI) requires delivery and installation of a Wallac
1420 Victor Multilabel Counter with a Wallac Lab Computer Pentium 100
w/16mb RAM and a HP Deskjet Printer from Wallac Inc., 9238 Gaither
Road, Giathersburg, MD 20877. The NCI Laboratory of Biochemistry
requires the equipment to allow investigators to read four major label
families in a single compact instrument; absorbance, fluorescence,
time-resolved fluorescence, and chemiluminescence. NCI's specific
requirements include continuous light source for fluorometric and
photometric measurements: 1) tungsten-halogen lamp, 75w, spectral range
350-700 nm, 2) rotatable filter wheel, provided with 8 filter positions
and standard high quality interference filters 405 nm, 450 nm, 490 nm,
355 nm, 485 nm. The detection unit for photometry must be a photodiode
with photometric range 400-700 nm. For fluorometry, luminometry, and
TR-fluorometry the detection unit must include photommultiplier tube,
R1527 and emission filter slide provided with 6 filter positions-615
nm, 642 nm, 545 nm, 572 nm, 535 nm, 460 nm. Each detection system must
be equipped to read 24, 48, 96 and 384 well microtitration plates. The
measurement time must be at least 50 seconds for 96 well plates. The
detection limit for fluorescein must be (10 fmol/well; europium
detection limit (10 amol/well; Eu-linearity 5 decades; measurement
range for photometry @ 405 nm 02-A; with accuracty <2% and precision
<0.5%. For luminometry the lower limit of detection must be less
than a total flux of 10E5 photons/s. The machine must be equipped with
in and out stackers for microtitrationn plates. The system requires a
Pentium (Intel) compatible PC with 16 Mbyte RAM, 24 Mbyte hard disc
space, CD-Rom, serial and parallel porst and high resolution SVGA
display and Windows '95, and a HP Deskjet printer. The Wallace 1420
Victor Multilable Counter system is the system that NCI had identified
that meets these requirements. This notice of intent is not a request
for competitive proposals.However, if any interested party believes
that it can meet the above reuirements, it may submit a statement of
capabilities. Information furnshed shall include sufficient detail to
determin if the equipment meets the foregoing requirements, and must be
received within 20 days of the publicaiton of this notice. See Note 22.
(0094) Loren Data Corp. http://www.ld.com (SYN# 0349 19970408\66-0017.SOL)