COMMERCE BUSINESS DAILY ISSUE OF JANUARY 21,1997 PSA#1765Commander, NAVSURFWARCENDIV Crane, Code 1164, Bldg. 64, 300 Highway
361, Crane, IN 47522-5011 66 -- TEMPERATURE CHAMBER SOL N00164-97-Q-0078 DUE 031097 POC Contact
point, Ms. Hilda Quillen, Procurement Technician, Code 1164EJ, Phone
812-854-2422, Fax 812-854-3465, e-mail hq6591@smtp.nwscc.sea06.navy.mil
The Navy requires manufacture of one each temperature chamber providing
a minimum temperature range from -40 degrees to +350 degrees
fahrenheit. Control within +/- 1.8 degrees must be provided, with
temperature stabilization within 35 minutes. Material of manufacture
for the interior is Type 304 stainless steel; minimum interior free
space shall be 5 feet wide by 5 feet deep by 5 feet high. Exterior
dimensions shall not exceed 7 feet wide by 11 feet deep by 6 feet high.
Non-explosive and non-ozone depleting refrigerant is required.
Electrical service available at the site for installation is 460 volt,
3-phase, 60 cycle. A microprocessor-based temperature
programmer/controller and a circular chart recorder shall be provided.
Solicitation requests may be made to the above contact point; e-mail
is preferred. All responsible sources may submit an offer which will be
considered by the Navy. (0016) Loren Data Corp. http://www.ld.com (SYN# 0289 19970121\66-0010.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|