COMMERCE BUSINESS DAILY ISSUE OF AUGUST 22,1996 PSA#1664US EPA, CONTRACTS MANAGEMENT DIV, CINCINNATI, OH, 45268-7001 66 -- QUADRUPOLE MASS SPECTROMETER SOL RFQ 96-110 POC Barbara A.
Curtiss Gas chromatograph specifications: shall be configured to
accommodate the hewlett packard 7673a autosampler for liquid sample
introduction. Must interface to any tekmar purge-and-trap, headspace or
desorption unit. Data system must be able to control all autosampler
injection and operation parameters and recall multiple sets of
parameters. System will have a dual injector block design and an
independently heated grob-type injector. The gc column oven will have
an operating temp range of -99 degree c to 400 degree c -99 degree c to
400 degree c with subambient option using liquid nitrogen as a coolant.
Must have multi-linear temperature programming capability with up to
eight segments in increments as low as 0.1 degree c/min. Will be able
to store and recall multiple sets of temperature profiles and
instructions with all control and readout functions perfformed via the
data system workstation. Must be able to be changed during a mass spec
data acquisition run. Interface must be a direct line-of-sight
connection and must allow the capillary column to terminate within one
millimeter of the ion volume of the mass spectrometer ion source. Must
be able to access and remove the interface within five minutes. Must
have a binary solvent delivery system. The system will be able to use
a minimun column size of 1.0 mm diameter. Provide a column flow range
of 0.01 to 10 ml/min using a maximum pump pressure of at least 5800 psi
and a flow rate precision of (0.3%. must have a variable wave length uv
detector. The system shall be supplied with a hplc divert valve and
operate either in automated or manual mode. The system shall be
supplied with software kit. Kit shall include chameleon nfs/x
networking software and include configuration instructions. The system
shall be supplied with a kit to permit the interfacing of a capillary
electrophoresis (ce) instrument to the api source of the mass
spectrometer. The interface shall be supplied with a modified esi
nozzle and all required tubing. Patent license shall be provided.
Vendor shall supply a 21'' color monitor and initial supply of spare
parts and consumables. Vendor shall supply a complete set of all
operator and maintenance manuals and a one-year warranty. Delivery
shall be no more than 180 days after award. All interested parties
should submit a request in writing to Barbara A. Curtiss at the above
address or by fax (513-366-2107) referencing RFQ 96-110. No telephone
request will be honored. (0233) Loren Data Corp. http://www.ld.com (SYN# 0401 19960821\66-0007.SOL)