|
COMMERCE BUSINESS DAILY ISSUE OF july 5,1996 PSA#1630US Env. Protection Agency, CMD-RTP MD-33 (NCCM-H), Research Triangle
Park, NC 27711 66 -- TIME-OF-FLIGHT MASS SPECTROMETER SOL D600462M1 POC Lenora
Hilliard, 919-541-4392/Janice F. White, 919-541-4392. The Environmental
Protection Agency (EPA) has a requirement for a Time-of-Flight Mass
Spectrometer. THIS IS A COMBINED SYNOPSIS/SOLICITATION FOR COMMERICIAL
ITEMS PREPARED IN ACCORDANCE WITH THE FORMAT IN SUBPART 12.6, AS
SUPPLEMENTED WITH ADDITIONAL INFORMATION INCLUDED IN THIS NOTICE. THIS
ANNOUNCEMENT CONSTITUTES THE ONLY SOLICITATION; PROPOSALS ARE BEING
REQUESTED AND A WRITTEN SOLICITATION WILL NOT BE ISSUED (with the
exception of the listing of an approved third party Agency approval by
the North Carolina Building Code Council which will be telefaxed upon
request). The solicitation document and incorporated provisions and
clauses are those in effect through Federal Acquisition Circular 37.
The solicitation number is D600462M1 and it is being issued as a
Request for Proposal (RFP). The contractor shall provide a
time-of-flight Mass Spectrometer and data analysis system. The
following minimum specifications are as follows: (1) the system must be
capable of acquiring, processing, and storing mass spectra continuously
at a rate of at least 500 spectra per second; (2) ability to produce a
mass resolution of 1 atomic mass unit (amu), or better, over a
scanning range of -1 1000 m/z; (3) The components of the ion launch and
the detection electronics shall allow spectral acquisition rates of
1000 spectra per second, or faster; (4) ability to sum or average as
few as two spectra launches and acquisitions in order to produce a scan
rate of 500 scans per second (or Hertz (Hz)); (5) The operator shall be
able to select the number of spectra acquisition events to be averaged
together, and the system shall be capable of providing scan rates that
span the range of at least 10 scans per seconds (100 milliseconds per
scan) to 500 scans per second (2 milliseconds per scan) - the selection
of such scan rate will depend on the researcher's need or time
resolution and signal-to-noise ration; (6) The data analysis hardware
and software shall be capable of resolving (deconvoluting) spectra of
individual species from a complex chromatogram, including effective and
accurate techniques for subtraction of the background; (7) The system
shall also include software, hardware, and computer systems, as needed,
to set-up, control, and trouble-shoot the system; to report the
observed mass spectral information, together with single-ion or
total-ion chromatograms; to search library spectra and report
''matches'' with the observed spectral data; and to produce a
quantitative chromatographic report. (This quantitative chromatographic
report shall include, as a minimum, the retention time, the total ion
(and/or specific single ion) count, the name of the chemical based on
a library search of mass spectra, a measure of the quality of the mass
spectral identification, and the concentration of the chemical, based
on the actual or estimated calibration data.); (8) Data acquisition,
data processing, and data storage shall be adequate to meet the scan
rate requirements and to store spectra from a one hour-long run; (9)
system must be capable of operating, collecting, and producing
identifiable and interpretable spectra at input flow rates of up to 5
ml of nitrogen per minute in order to handle the interfaces that are
envisioned for the research efforts; (10) the system shall operate at
an ion launch and detection rate of at least 1000 scans per second
(this is required because the system will be used with fast
chromatographic systems that will produce peaks as narrow as 10
milliseconds in bandwidth; (11) the internal launch rate of individual
ion groups and the electronic data acquisition components must be
capable of producing at least 1000 scans per second; (12) The system
shall be capable of a storage rate of at least 500 scans per second,
however, a range of spectra storage rates, selectable by the operator,
covering at least the range of 10 scans per second to 500 scans per
second shall be available; (13) the vacuum system for the
time-of-flight mass spectrometer shall be capable of accepting,
operating, and collecting resolved mass spectral data with a continuous
input of carrier gas typical of that used in fast gas chromatographic
systems; (14) the system must be capable of accomodating an input of at
lease 5 milliliters per minute of nitrogen carrier gas, while
maintaining ion source pressure at an appropriate operating level.
(Note: the volume of 5 ml is at 1 atmosphere pressure and a temperature
of 25 degrees C.); (15) the system must be capable of measuring with a
signal-to-noise ratio of 7 or better, typical hydrocarbnon pollutants
in abmient air (e.g., benzene, toluene hydrocarbon, the xylenes,
n-hexane, n-octane) in ambient air at gas chromatographic (v/v) of 250
parts per trillion in a typical fast gas chromatographic run(e.g,. 1-3
minutes total run time, with peak widths of 0.1-0.5 second; (16)
ability to measure carbon tetrachloride with a signal-to-noise of 7 or
better in ambient air at its geochemical background concentration of
approximately 125 ppt (v/v); (17) system must be capable of
distinguishing among coeluting compounds since no chromatograph is
capable of resolving all of the analytes in a typical ambient air
sample; (18) system shall provide an effective, demonstrable
deconvolution algorithm as past of its data processing software and the
algorith must me capable of resolving, identifying, and quantifying the
concentrations of individual compounds, including volatile organic
compounds identified as hazardous air pollutants in 42 U.S. Code 7112
(Section 112 of the Clean Air Act as amended) and other species of
interest; (19) system must have a mass scanning range of 1 to 1000 mass
units (m/z); (19) The mass spectrometer, excluding computer equipment
and other peripheral equipment (e.g., roughing pump, or a dedicated
printer or monitor) shall weight no more than 400 pounds, and shall
have a ''footprint'' no larger than 3 feet by 4 feet; (20) The mass
spectrometer shall not require more than 20 ampere service at 120 volts
AC (60 Hz or 225 V ac (60 Hz); (21) The mass spectrometer must be
capable of operating across the range of 50 degrees to 90 degrees
fahrenheit, with relative humidities of 20% to 80%; (22) The equipment
must be underwriters laboratory approved, or it shall be listed and
labeled by a third party Agency approved by the North Carolina Building
Code Council. The contractor shall deliver the instrument to the
Environmental Protection Agency, Research Triangle Park, N.C. The
contractor shall install the equipment and demonstrate that it meets
the required specifications. Additionally, the contractor shall conduct
training for up to five (5) operators. The Contractor shall offer to
the Government at least the same warranty terms, including offers of
extended warranties, offered to the general public in customary
commercial practice. Brochures or manufacturers specifications must be
included with your bid/offer. Deliver is required within six (6)
months after contract award. Contractor shall notify the Project
Officer at least 2 weeks prior to shipping the mass spectrometer.
Shipment shall be FOB Destinatioin to U.S. Environmental Protection
Agency, Shipping and Receiving, Page Road and 1-40, Research Triangle
Park, NC 27709. The following FAR provisions apply to this
solicitation: 52.212-1., Evaluation -Instructions to
Offerors-Commerical; 52.212-2, Evaluation-Commericial Items. Evaluation
criteria to be included in paragraph (a) of provision 52.212-2 are as
follows: (1) technical capability of the item offered to meet the
Government requirements; (ii) past performance; (iii) price, in
decreasing order of importance. Technical and past past performance,
when combined, are significantly more important than price. All
offerors are to include with their offers a completed copy of provision
52.212-3. Offeror Representations and Certifications-Commericial Items.
The following FAR clause applies to this acquisisition: 52.212-4,
Contract Terms and Conditions-Commericial Items. The following
additional FAR clauses, which are cited in clauses 52.212-5 are
applicable to the acquisition: 52.203-6, Restrictions on Subcontractor
Sales to the Government, with Alternate I; 52.203-10., Price or Fee
Adjustment for Illegal or Improper Activity; 52.219-8, Utilization of
Small Business Concerns and Small Disadvantaged Business Concerns;
52.222-26, Equal Opportunity, 52.222-35, Affirmative Action for
Hanficapped Workers; 52.222-37, Employment Reports on Special Disabled
Veterans and Veterans of the Vietnam Era; 52.225-3, Buy American
Act-Supplies; 52.225-18. European Community Sanctions for End Products;
52.225-21, Buy American Act- North American Free Trade Implementation
Act-Balance of Payment Program, Clause EP 239-103, Acquisition of
Microcomputers Imcluding Personal Computers, Monitors, and Printers
(Energy Star Requirements) is also applicable to this requirement.
Offerors must submit all technical questions concerning this
solicitation in writing to the Contract Specialist. All bid/offerors
should be submitted to Lenora Hilliard, Contract Specialist, U.S.
Environmental Protection Agency, MD-33, Research Triangle Park, NC
27711. Courier Delivery Address is U.S. Environmental Protection
Agency, Receptionist-Administration Bldg. Lobby, Attn: Lenora Hilliard,
79 Alexander Drive, Research Triangle Park, NC 27709. All offers are
due by July 17, 1996, 4:30 p.m. EST. Offers/bids received after the
exact time specified will not be considered. No telephonic or faxed
requests will be honored (0184) Loren Data Corp. http://www.ld.com (SYN# 0411 19960703\66-0001.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|