COMMERCE BUSINESS DAILY ISSUE OF JULY 2,1996 PSA#1628Operational Contracting Division, 1940 Allbrook Dr, Ste 3 Bldg 1, WPAFB
OH 45433-5309 66 -- DIRECTIONAL SOLIDIFICATION UNIT SOL F33601-96-R-0181 POC For
copy, ASC/PKWO, Attn: Bldg 260, 1940 Allbrook Dr Ste 3, WPAFB OH
45433-5309, POC: C Marion, 513-257-5197, Fax No 513-257-3926. A
solicitaion is not available at this time. Sources are sought for a
Directional Solidification Unit. A vertical, directional Solidification
(D.S.) unit based on the induction float-zone method is required for
processing titanium alloys and titanium aluminides ona laboratory
scale. The unit should be capable of producing single crystals or
aligned, columnar grains. The system should allow experimental
characterization of solidification microstructures obtained at
different values of temperature gradients and solidification rates
under both transient and steady-state conditions. TECHNICAL
SPECIFICATIONS - MELTING AND SOLIDIFICATION CHAMBER: Melting and
directional solidification shall be conducted inside a chamber which is
capable of operating in vacuum better than 10(3) torr as well as
positive pressure up to 10 psi argon. The chamber should have a minimum
internal diameter of 300 mm (12 in.) and a height of 1000 mm (40 in.).
The chamber should be made of type 304 stainless steel with double
-wall construction suitable for effective water cooling. The chamber
should be equipped with a water-cooled door of at least 600 mm (24 in.)
height. The door should be provided with a convenient and quick
mechanism for opening and closing of the chamber and appropriate
sealing mechanism to achieve vacuum/pressure specified above. The
chamber should also have an operator viewpoint made of clear quartz
with a minimum size of 100 mm (4 in.). In addition, the chamber should
have three mounting ports, one for R.F. feed-through, one for
pyrometer and the third for unspecified use. Standard ports and
connections for the vacuum line, over-pressure valve and argon gas flow
should also be incorporated. INDUCTION HEATING SYSTEM: The user will
provide the induction power supply system, including the working coil
and the flux concentrator plate. However, the D.S. unit should include
a suitable arrangement involving bellows and coaxial feed-through in
order to align the melt stock and the coil. FRAME AND SUPPORTS: The
D.S. unit should be equipped with sturdy support structures for
supporting the chamber and the feeding system. It should also include
appropriate leveling mechanisms. WITHDRAWAL SYSTEM: The withdrawal
system should be capable of pulling the solidified rods, with diameters
ranging from 10-25 mm (0.375-1.0 in.) from the melt zone. The
withdrawal shaft should be made of 304 stainless steel and have an
internal water cooling arrangement. The displacement of the shaft shall
be accomplished by a precision ball screw and nut assembly driven by a
servomotor or a stepper motor. Minimum Stroke Length: 300 mm (12 in.).
Withdrawal Lift Speed: 0.05-15 mm/min, (002-0.6 in./min.) (with an
accuracy better than 1%). Shaft Rotation Speed: 1-30 rpm. Shaft Seal:
Sealing between the chamber and shaft should be effective under vacuum
and positive pressure. FEED SYSTEM: The feed system supplies the melt
stock with the diameters ranging from 10 to 25 mm (0.375-1.0 in.) to
the molten zone. The shaft should be made of 304 stainless steel and be
water cooled. The displacement of the shaft shall be accomplished
through a precision ball screw and nut assembly. Minimum Stroke Length:
300 mm (12 in.). Feed Rate: 0.05-15 mm/min. (0.002-0.6 in./min.) (with
an accuracy of better than 1%). Shaft Rotation Rate: 1-30 rpm. Shaft
Seal: Seal between the shaft and the chamber should be effective under
vacuum and positive pressure. Alignment: A ball slide and rack
assembly capable of adjustments in both x and y dierections up to 6 mm
(0.25 in.) is to be provided for alignment of top feed rod and the
bottom pull rod. VACUUM/INERT GAS SYSTEM: This should consist of a 51
mm (2.0 in.) vacuum manifold, a main vacuum valve 51 mm (2.0 in.) with
an actuator and an over-pressure valve. Suitable transducers and gages
for displaying the chamber vacuum and positive pressue up to 7 kPa (10
psi) shall be incorporated. A suitable system for controlled
admittance of argon gas and air into the chamber shall be incorporated.
Vacuum pump(s) for this system will be supplied by the user. WATER
COOLING SYTEM: A suitable water cooling arrangement with inlet and
outlet mainifolds, pressure gages, valves, temperature indicatiors and
pressure relief valves shall be incorporated. OPTIONAL ITEM - The
supplier may provide the following parts as optional items: CONTROL
SYSTEM: A simple yet versatile control system for the D.S. unit is
required. It should include both hardware and software required to
drive the motors for the withdrawal and feed systems, a feedback
process control and data acquisition system, including power control
for the induction heater and safety interlocks for water, argorr and
vacuum lines. The data acquisition system should be capable of
collection, display, storage and analysis of the following
signals/parameters: power output, melt temperature, feed rate,
withdrawal rate and axial temperature profile. The Standard Industrial
code is 3339. Therefore, the small business size standard is 750
employees. Interested sources are required to submit proof of
capability to ASC/PKWOCB, 1940 Allbrook drive, Ste 3, WPAFB OH
45433-6503, not later than 12 Jul 96. Also state your business size.
Direct all requests for solicitations and routine communication
concerning this acquisition to Cindy Marion, Contracting Specialist,
(513) 257-5197. An Ombudsman has been established for this acquisition.
The only purpose of the Ombudsman is to receive and communicate serious
concerns from potential offerors when an offeror prefers not to use
established channels to communicate his/her concern during the proposal
development phase of this acquisition. Potential offerors should use
established channels to request information, pose questions, and voice
concerns before resorting to use of the Ombudsman. Potential offerors
are invited to contact ASC's Ombudsman, Col Michael J. Corcoran,
ASC/IG, Bldg 14, 1865 Fourth St, Ste 12, Wright-Patterson AFB OH
5433-7126, at (513) 255-0600, with serious concerns only. (0180) Loren Data Corp. http://www.ld.com (SYN# 0404 19960701\66-0019.SOL)