|
COMMERCE BUSINESS DAILY ISSUE OF JUNE 17,1996 PSA#1617NASA Lewis Research Center, 21000 Brookpark Road, Cleveland, Ohio
44135 66 -- SCANNING ELECTRON MICROSCOPE SYSTEM - PART 1 SOL SFO3-051357 POC
Glen M. Williams, Contracting Officer, who may be contacted at (216)
433-2885, Mail Stop 500-306 This is a combined synopsis/SFO for
commercial items prepared in accordance with FAR Subpart 12.6, as
supplemented with additional information included in this notice. This
announcement constitutes the only solicitation; offers are being
requested and a written solicitation will not be issued. This SFO and
incorporated provisions and clauses are those in effect through Federal
Acquisition Circular 90-37. The Standard Industrial Classification
(SIC) Code for this acquisition is 3826. The Small Business Size
Standard is 500 employees. Under the Defense Priorities and Allocations
System (DPAS)(CFR 15 Part 350), the rating assigned to this acquisition
is DO-C9. This procurement is being conducted under the NASA Midrange
Pilot Test Program approved by the Office of Federal Procurement Policy
on April 16, 1993. The specifications for the items are as follows:
Specifications for Transmission Electron Microscope,
Scanning/Transmission 200KV. GENERAL-The Contractor shall furnish one
high resolution analytical scanning transmission electron microscope
with electron imaging filter and x-ray energy dispersive spectroscopy
system which shall include the following items; ITEM 1. The Contractor
shall provide a scanning/transmission electron microscope (STEM) which
shall have the following features and capabilities; Item 1A. Vacuum
System 1. Specimen area, electron source chamber, and microscope column
down to the final projector lens shall be pumped by an oil-free pump.
Vacuum surfaces and components in the vicinity of the electron beam
shall not be subject to degassing or degradation under electron or
x-ray radiation. 2. A liquid nitrogen cold trap around the specimen
shall be included. Cold trap shall be compatible with detectors and
specimen holders as specified below. 3. A vacuum of less than 3 x
10E-5Pa shall be maintained in the specimen area, with partial
pressures of hydrocarbons less than 4 x 10E-8 Pa and of water vapor
less than 8 x 10E-6 Pa. 4. A pressure difference of 104 shall be
maintained between plate camera/projection chamber and microscope
column via a differential pumping aperture. This aperture shall not
degrade the optical performance of the instrument. Frequent camera
changes shall be permitted without degrading the microscope column
vacuum. This aperture shall provide a fixed crossover plane for fully
focused energy loss spectrometry. 5. The vacuum system shall be
protected from damage in the event of power and/or water failure Item
1B. Specimen Airlock / Stage 1. Specimen holder insertion shall be
through an automatic pre-pumped airlock with high voltage and filament
heating remaining on during specimen exchange. 2. Airlock and stage
shall be compatible with Government owned specimen holders currently
used on a Philips 400T. Inventory includes; one single tilt (Philips
PW6596/00), one low background single tilt (Philips PW6597/00), two low
background double tilt (Philips PW6595/00), one double tilt liquid
nitrogen (Gatan 636), one single tilt heating (Gatan 628TA), one low
background tilt/rotate (Gatan 650 Mark I), and one single tilt multiple
specimen holder (Philips PW6598/02). If the Contractor's system can not
be made compatable with the above Government inventory, the Contractor
shall furnish the above stated items compatable with their System. 3.
Contractor shall supply a low background single tilt holder and a low
background double tilt holder. Both holders shall have mechanisms
capable of holding brittle specimens without damage. 4. Specimen holder
drift on a temperature stabilized holder shall be less than one (1)
nanometer per minute. 5. Specimen stage shall be a fully motorized,
5-axis (X, Y, Z, alpha, beta), side entry design with alpha tilt range
of plus or minus seventy (70) degrees. Control of the five axes shall
be by both internal microprocessor and external computer. 6. Internal
microprocessor control shall display X, Y, and Z to 0.1 micrometer and
alpha and beta to 0.1 degree and provide storage and recall of a
minimum of 25 sets of stage positions as defined by the five axes. 7.
X and Y movements shall maintain vertical and horizontal directions on
the main viewing screen and TV monitor at all magnifications. The
relative speed of image movement on the main screen, binocular, and TV
monitor shall be operator selectable and remain constant at all
magnifications. 8. The specimen stage shall have a direct position
measurement system that is free of backlash, orthogonal hysteresis, and
run-on. 9. Fine X and Y adjustments shall allow specimen movements as
low as 2 nanometers. Reproducibility and absolute measurement of X/Y
positioning shall be less than 0.5 micrometer and angular positioning
shall be less than 0.5 degrees. 10. The internal microprocessor control
shall include an alpha wobbler for fine adjustment of Z to the
eucentric position. 11. Software shall be provided to allow an Apple
Macintosh computer to control the stage to perform automatic grid
searches, image montages, and storage and recall of stage's five axes'
positions. Item 1C. Electron Illumination Source 1. Range of
accelerating voltages shall be to a maximum of 200KV with continuous
control in nine (9) operator selectable step sizes between 50 volts and
10 KV per step. 2. Use of tungsten and/or lanthanum hexiboride emitters
shall be allowed without modifying the vacuum system. 3. Microprocessor
control shall prevent overheating of the emitter and provide automatic
warm-up procedure for lanthanum hexiboride emitters to prevent
premature failure. 4. All column alignments shall be carried out using
electromagnetic beam deflectors. 5. Contractor shall provide two (2)
wehnelts, both capable of using lanthanum hexiboride emitters. 6.
Contractor shall provide two (2) lanthanum hexiboride emitters. Item
1D. Illumination System 1. The condenser lens aperture holder shall be
capable of accepting four (4) standard or low background 3 mm
apertures which can be selected by a click-stop mechanism. 2.
Contractor shall include three (3) sets of four (4) low background
apertures of standard sizes. 3. Microprocessor shall control the
illumination system such that the intensity remains constant with
changing magnification, or the maximum intensity is limited, or allow
unrestricted control. 4. Illumination system shall be capable of
achieving a minimum TEM mode spot size of 2 nm or less with a lanthanum
hexiboride emitter with the specimen in the eucentric position. 5.
Illumination system shall be constructed to minimize generation of
superious x-rays. Item 1E. Objective Lens 1. Objective lens design
shall allow the specimen to remain at the eucentric position for all
operating modes. 2. Objective lens shall have a Scherzer resolution of
less than or equal to 0.27 nm at 200 KV and lattice resolution of less
than or equal to 0.14 nm at 200 KV. 3. Electrical stability shall be
less than one (1) part per million per minute at 200 KV. 4. Minimum
focus step shall be less than or equal to 3 nm. 5. Microprocessor shall
automatically adjust focus step size with magnification with manual
override possible. 6. Objective lens shall be at constant excitation to
at least 600,000X to minimize thermal changes in the specimen area and
to eliminate the need to adjust the objective stigmator. 7. Objective
aperture mechanism shall hold a minimum of four different size
apertures, selectable by a click-stop mechanism, and allow full
retraction. 8. Contractor shall include three (3) sets of four (4)
platinum apertures of standard sizes. 9. Microprocessor control shall
provide ability to select one of five previously chosen under focus
conditions or to select Scherzer de-focus. (0165) Loren Data Corp. http://www.ld.com (SYN# 0327 19960614\66-0002.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|