COMMERCE BUSINESS DAILY ISSUE OF APRIL 26,1995 PSA#1333Directorate of Contracting, Fitzsimons Army Medical Center, ATTN:
MCHG-DCC, Aurora, CO 80045-5001 65 -- MEDICAL WASTER STERILIZER SYSTEM SOL DADA03-95-R-0018 DUE 061495
POC Jackie Scripture, Contract Specialist, (303)361-3315 or Emily J.
Mathis, Contracting Officer, (303)361-3315. Fitzsimons Army Medical
Center has a requirement for one (1) medical waste sterilizer system,
an Ongard Medical WasteClave Model with a single door, or equal.
Required salient features include, but are not limited to the
following: 1) Must be a steam sterilizer system with stainless steel
chambers, and stainless steel or other equally durable material for the
jacketing; 2) The pressure vessel must have inside dimensions of at
least 36'' x 48'' x 72'', must be loaded and unloaded from a single
door, must be certified to ASME Code, 316L Construction, and must fit
within the available space at Fitzsimons; 3) The operational system
shall include a high performance vacuum pump system with valves being
air operated and all interlocks National Board certified; 4) The
controller shall be programmable with easily read alpha and numerical
printer and display. 5) The display and printer shall be capable of
printing the following information: cycling time, key transition
points, cycle summary and temperature log; 6) The system must be able
to operate for a minimum of an eight (8) hour period without
electricity if the power supply is temporarily cut off (may offer a gas
generator or UPS); 7) Must have a compressor capable of 1/4'' NPT, 125
PSI Min, 7 SCFM; 8) The loading system shall have at least one inner
cart and have an external transfer carriage system; 9) System shall
have a one year warranty on parts and labor and at least a five year
warranty on the pressure vessel; 10) All services and
supplies/equipment for installation and any necessary testing shall be
provided by the Contractor and result in a fully functioning system
upon completion of installation. Two sets of technical data consisting
of operating instructions, maintenance and calibration procedures,
schematics A/O wiring diagrams and a repair parts list shall also be
provided. The system must be capable of using the utilities available
on site, which are: Power: 3-Phase -208V, Single Phase -110V; Steam: 65
PSI Average, 2 inch line; and Water: Capable of a peak of 20 GPM, 2
inch line. Delivery is to be FOB DESTINATION, 120 days ARO, with
installation to be completed within 30 days of receipt. The
solicitation is to be issued o/a May 15, 1995 with offers due 30 days
after actual issuance date. All responsible sources may submit an offer
which will be considered. Faxed requests will be accepted at
(303)361-4208. All requests for solicitations must be submitted in
writing. (0114) Loren Data Corp. http://www.ld.com (SYN# 0388 19950425\65-0001.SOL)
65 - Medical, Dental and Veterinary Equipment and Supplies Index Page
|
|