COMMERCE BUSINESS DAILY ISSUE OF FEBRUARY 16,1995 PSA#1285NIH, National Heart, Lung, & Blood Institute, Procurement Section,
Contracts Operations Branch, Westwood Building, Room 836, 5333 Westbard
Avenue, Bethesda MD 20892-9070 66 -- ATOMIC ABSORPTION SPECTROMETER SOL NHLBI-HL-P-95-102 POC
Kathleen J. Van Steelant, Contract Specialist, (301) 594-7874. It is
the intent of the National Heart, Lung, and Blood Institute (NHLBI) to
acquire a Atomic Absorption Spectrometer for installation in the
Laboratory of Biochemistry in Bethesda, Maryland. The instrument must
meet the following minimum specifications: 1) The atomic absorption
spectrometer should be a furnace system, fully automatic and able to
run up to 4 elements automatically and sequentially. The graphite
furnace must be heated transversely to allow uniform heating across the
whole length of the graphite tube, which shall have an integrated L'vov
platform that is touching only one side of the graphite tube; 2) The
background correction for the furnace must be accomplished with
Longitudinal Zeeman, so no polarizer is needed in the optical path; 3)
A monochromator with a focal length of at least 260 nm and a
reciprocal linear dispersion of 1.6nm/mm; 4) A furnace autosampler must
be provided that offers at least 81 sampling positions; 5) A
electrodeless power supply must be provided that has push button
ignition of the EDL lamps; 6) A system controlled fume extraction
system must be provided with the graphite furnace to automatically
remove all vapors and materials expelled by the furnace (smoke, bad
odors, corrosive fumes) during drying and sample pretreatment stages,
to help reduce the time needed for these stages; 7) A water circulating
cooling system must be provided that is controlled by the system and
automatically turns off by the system after an automatic run; 8) A
single controlling computer (PC type with at least 8 megabyte RAM, 200
megabyte hard disk, and 80486 CPU) with printer must be provided; and
9) The following hollow cathode lamps elements must be provided:
Copper, Zinc, Iron, and Nickel. A Selenium electrodeless discharge lamp
must also be provided. The offeror shall provide training on the
instrument for two (2) operators who will be designated by the NHLBI
Project Officer. A firm fixed-price type contract will be negotiated.
Delivery and installation of the instrument is anticipated within
thirty (30) calendar days after the contract award. Only a fully
installed and functional instrument will be accepted. Tests will be
performed to ascertain that the instrument is functional before final
acceptance. It is anticipated that Request for Proposal (RFP)
NHLBI-HL-P-95-102 will be issued on or about March 1, 1995. Copies of
the RFP may be obtained by writing to the National Heart, Lung, and
Blood Institute, Contracts Operations Branch, Procurement Section,
Westwood Building, Room 836, 5333 Westbard Avenue, Bethesda, Maryland
20892, Attention: Kathleen J. Van Steelant, Contract Specialist.
(0045) Loren Data Corp. http://www.ld.com (SYN# 0207 19950215\66-0001.SOL)
66 - Instruments and Laboratory Equipment Index Page
|
|