66 -- Gas Chromatograph

Notice Date
5/30/2023 12:06:25 PM
 
Notice Type
Combined Synopsis/Solicitation
 
NAICS
334516 — Analytical Laboratory Instrument Manufacturing
 
Contracting Office
DEPT OF COMMERCE NIST GAITHERSBURG MD 20899 USA
 
ZIP Code
20899
 
Solicitation Number
1333ND23QNB640306
 
Response Due
6/7/2023 8:00:00 AM
 
Archive Date
06/22/2023
 
Point of Contact
Tish Walker
 
E-Mail Address
latish.walker@nist.gov
(latish.walker@nist.gov)
 
Small Business Set-Aside
SBA Total Small Business Set-Aside (FAR 19.5)
 
Description
The purpose of Amendment I is to: Incorporate amended Statement of Requirements document; & Respond to industry inquiries. All other terms and conditions remain unchanged, as a result of this amendment.� Question - The SOW specifies FID and ECD, but the analysis requirement specifies TCD to detect down to ppb that is not practically achieved. Does your scientist need TCD for oxygen/air/methane and ECD for Matrious oxides detection? The desired detectors are a flame ionization detector (FID) and electron capture detector (ECD). This error appears in both �I. Specifications� and �VI. Acceptance criteria�. The ECD will be used for nitrous oxide at, or around, a nominal 300 ppb (nmol/mol). Second question is about 10ml gas sampling loop.� This is very high for gas chromatography technique.� Is the 10ml a typo of 1.0ml? In some cases, trace analyses, the sample loop will need to be 10 mL. The third question is about sample injector. Is this sample inlet a packed inlet, split/splitless type? Does your scientist need an autosampler for liquid sample injection or for large volume gas sample? An autosampler for liquid samples is not needed here. This GC will be used for analyses of high-pressure gas samples delivered at a controlled pressure between 20 � 40 psig. The inlet will need to support these analyses performed with packed columns. The RFQ� specifies a GC instrument with both Flame Ionization detection and electron capture detection but the system specifications only Flame Ionization and Thermal conductivity detection.� Is this for a single GC and if so which detectors are you requiring? This is for a single GC with two detectors, a thermal conductivity detector and electron capture detector. COMBINED SYNOPSIS/SOLICITATION 1333ND23QNB640306 Gas Chromatograph This is a combined synopsis/solicitation for commercial items prepared in accordance with the format in FAR Subpart 12.6 � Streamlined Procedures for Evaluation and Solicitation for Commercial Items, as a supplemental with additional information included in this notice. This announcement constitutes the only solicitation; quotations are being requested, and a separate written solicitation document will not be issued. This solicitation is being issued under the authority of FAR Part 13, Simplified Acquisition Procedures. This is a solicitation is a Request for Quotation (RFQ). The solicitation document and incorporated provisions and clauses are those in effect through Federal Acquisition Circular (FAC) 2023-02 effective 3/16/2023. The associated North America Industrial Classification System (NAICS) code for this procurement is 334516 � Analytical Laboratory Instrument Manufacturing with a Small Business size standard of 1,000 employees. This acquisition will be solicited as a Total Small Business Set-Aside.
 
Web Link
SAM.gov Permalink
(https://sam.gov/opp/e033b986243d41149c5ac6df883aa20a/view)
 
Place of Performance
Address: Gaithersburg, MD 20899, USA
Zip Code: 20899
Country: USA
 
Record
SN06698395-F 20230601/230530230122 (samdaily.us)
 
Source
SAM.gov Link to This Notice
(may not be valid after Archive Date)

FSG Index  |  This Issue's Index  |  Today's SAM Daily Index Page |
ECGrid: EDI VAN Interconnect ECGridOS: EDI Web Services Interconnect API Government Data Publications CBDDisk Subscribers
 Privacy Policy  Jenny in Wanderland!  © 1994-2024, Loren Data Corp.