J -- BASE PLUS 4 STERRAD NX STERILIZER - Attachment

Notice Date
11/2/2015
 
Notice Type
Attachment
 
NAICS
811219 — Other Electronic and Precision Equipment Repair and Maintenance
 
Contracting Office
Department of Veterans Affairs;Network Contracting Office 9;1639 Medical Center Parkway;Suite 400;Murfreesboro TN 37129
 
ZIP Code
37129
 
Solicitation Number
VA24916Q0041
 
Response Due
11/9/2015
 
Archive Date
12/9/2015
 
Point of Contact
PHON PHASAVATH
 
E-Mail Address
ESBORO
 
Small Business Set-Aside
Total Small Business
 
Description
Statement of Work On-Site Preventive Maintenance of Sterrad NX Sterilizers 1. EQUIPMENT DESCRIPTION: Unit STERRAD NX Serial Number 33090295 Murfreesboro Campus Unit STERRAD NX Serial Number 33090293 Nashville Campus Unit STERRAD(R) 100NX(TM) 1-DR Serial Number 1041120448 Nashville Campus 2. On-Site service to perform corrective maintenance and semi-annual preventive maintenance (PM1 and PM2) on the Sterrad NX sterilizers to include all parts, labor and travel. The PM2 service includes all work required during the PM1 service and additional items included only in the PM2, as follows: Preventive Maintenance 1 (PM1): a.Change vacuum pump oil b.Replace the exhaust filter c.Clean the vaporizer bowl d.Drain water from the air storage tank e.Drain pressure regulator moisture trap f.Clean air filters/panel assemblies g.Inspect/Clean electrode (replace after 2 years) h.Inspect/Clean door, inside chamber i.Product verification j.Update PM log and reset PM interval Preventive Maintenance 2 (PM2): a.Perform Baratron Zero Shift test b.Replace H2O2 lamp c.Replace catalytic converter d.Clean platen surface e.Inspect/Replace door seal o-ring f.Replace shelves, stops, rails, and electrode spacers 3. Vendor to provide a written service report to include travel time, labor time and detail parts used with part# and description. Service report must be signed by an authorized VA employee from SPS or Biomedical Engineering. 4. Vendor to contact Biomedical Engineering to schedule repair, when he/she arrives at the VA to commence work and upon completion of work to demonstrate the unit is operational prior to departure from the VA.
 
Web Link
FBO.gov Permalink
(https://www.fbo.gov/spg/VA/NaVAMC/VAMCCO80220/VA24916Q0041/listing.html)
 
Document(s)
Attachment
 
File Name: VA249-16-Q-0041 VA249-16-Q-0041.docx (https://www.vendorportal.ecms.va.gov/FBODocumentServer/DocumentServer.aspx?DocumentId=2399567&FileName=VA249-16-Q-0041-000.docx)
Link: https://www.vendorportal.ecms.va.gov/FBODocumentServer/DocumentServer.aspx?DocumentId=2399567&FileName=VA249-16-Q-0041-000.docx

 
Note: If links are broken, refer to Point of Contact above or contact the FBO Help Desk at 877-472-3779.
 
Place of Performance
Address: THE DEPARTMENT OF VETERANS AFFAIRS;1310 24TH AVENUE SOUTH;NASHVILLE TN
Zip Code: 37212
 
Record
SN03935385-W 20151104/151102234027-8da2c0c92cd9e874a611f5270e714253 (fbodaily.com)
 
Source
FedBizOpps Link to This Notice
(may not be valid after Archive Date)

FSG Index  |  This Issue's Index  |  Today's FBO Daily Index Page |
ECGrid: EDI VAN Interconnect ECGridOS: EDI Web Services Interconnect API Government Data Publications CBDDisk Subscribers
 Privacy Policy  Jenny in Wanderland!  © 1994-2024, Loren Data Corp.