- Description
- In response to questions: QUESTION: In regards to the specifications for NB685010-15-01406 - Gas Analyzer Pumping Cart. Please clarify on one specification which is unclear: "Fore-vac max for N2: > 10 Torr". If it read < 10 Torr this would make sense, but > 10 Torr leave sampling pressure requirement open ended. ANSWER: The specification is only on the turbo-pump, not the entire system. To be more explicit, the turbo-pump has a maximum foreline pressure that it can tolerate in order to operate. Our specification is that the maximum tolerable foreline pressure is at least 10 torr. Many commercially available turbo-pumps easily meet or exceed this specification.
- Web Link
-
FBO.gov Permalink
(https://www.fbo.gov/spg/DOC/NIST/AcAsD/NB685010-15-01406/listing.html)
- Record
- SN03665490-W 20150314/150312234530-26162644202dda1c80fc4ec243304347 (fbodaily.com)
- Source
-
FedBizOpps Link to This Notice
(may not be valid after Archive Date)
| FSG Index | This Issue's Index | Today's FBO Daily Index Page |