- AMENDMENT 2: THE PURPOSE OF THIS AMENDMENT IS TO ANSWER QUESTIONS AND AMEND SPECIFICATION NUMBER 8. QUESTION 1: IS THE 15 MINUTE COOLDOWN TO 10K A STRICT REQUIREMENT? ANSWER 1: YES IT IS. QUESTION 2: IS THE AUTOMATION A STRICT REQUIREMENT? ANSWER 2: THE FULLY AUTOMATED TEMPERATURE CONTROL OVER THE WHOLE TEMPERATURE RANGE VIA COMPUTER INTERFACE IS A STRICT REUQIREMENT, HOWEVER, AUTOMATION FOR THE PUMP IS NOT. SPECIFICATION 8 IS AMENDED BELOW. QUESTION 3: DOES THE CRYOSTAT NEED A CONSTANT SUPPLY OF LN2 AND LIQUID HELIUM: ANSWER 3: THE CRYOSTAT WILL USE A CONSTANT SUPPLY OF CRYOGEN TO RUN EXPERIMENTS AT A DESIRED TEMPERATURE. IT WUILL USE EITHER LN2 (>77K) ONLY OR L HE (>4 K) FOR MEASUREMENTS. THE CRYOGEN IS SEPARATE AND THE CRYOGEN SUPPLY IS NOT PART OF THIS SOLICITATION. QUESTION 4: ARE THERE TWO SEPARATE HOSES FOR THE TWO CRYOGENS (HELIUM AND LN2)? ANSWER 4: NO QUESTION 5: IS A TRIPLE WALLED HOSE REQUIRED IN WHICH THE LIQUID NITROGEN IS KEEPING THE LIQUID HELIUM SUB-COOLED? ANSWER 5: NO SPECIFICATION 8 IS HEREBY AMENDED AS FOLLOWS: FROM: COOL-DOWN AND PUMP-OUT MUST BE FULLY AUTOMATED AND CONTROLLED VIA COMPUTER INTERFACE. TO: COOL-DOWN MUST BE FULLY AUTOMATED AND CONTROLLED VIA COMPUTER INTERFACE. PUMP-OUT MAY BE MANUALLY CONTROLLED OR FULLY AUTOMATED. ALL OTHER TERMS REMAIN UNCHANGED.
- Web Link
-
FBO.gov Permalink
(https://www.fbo.gov/spg/DOC/NIST/AcAsD/SB1341-12-RQ-0355/listing.html)
- Place of Performance
- Address: 100 BUREAU DRIVE, GAITHERSBURG, Maryland, 20899, United States
- Zip Code: 20899
- Zip Code: 20899
- Record
- SN02798798-W 20120712/120711000051-4e6861abbb45f2e3f3bff89d446e2c0a (fbodaily.com)
- Source
-
FedBizOpps Link to This Notice
(may not be valid after Archive Date)
| FSG Index | This Issue's Index | Today's FBO Daily Index Page |